AvCount Air provides reliable results for determining the particle concentration in air, for applications such as clean rooms, laboratories and the monitoring of air cleanliness in hospitals.
• Simple operation
• 3 measuring channels
• 6 particle size classes
• Programmable via PC
With a simple test procedure and user friendly features, operator skill or expertise is kept to a minimum.
AvCount Air can hold up to 3 test methods with custom methods created on a PC before being uploaded.
Test progress and results for up to 6 size bands are displayed as cumulative particles/ml.
Features
- Cumulative Particles/ml
- 0.21 µm to 10 µm
- Simple operation
- High visibility screen
- Stand-alone or PC controlled
- Programmable via PC
Additional Info
Results AnalysisThe AvCount Air can be controlled from a personal computer via the USB port. This allows the configuration of test parameters, test control and download of data. It also allows operation of a single device or multiple instruments simultaneously. The measured data is displayed and can be retrieved from the data memory later. An Excel template is available for exporting the data.
Calibration & Verification
Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated Master AvCount Air. Calibration can be carried out at a user site or at our authorised regional laboratories.
Particle Size Range |
0.21µm to 10µm |
Number of Measuring Channels |
3 (16 via USB) |
Particle Concentration |
250,000 P/sec |
Flow Rate |
28.3 l/min |
Ambient Temperature Range |
5 to 40 °C |
Air Temperature Range |
0 to 70 °C |
Results |
600 internal, unlimited via USB |
Software |
Protrend via USB |
Connectivity |
Via ethernet |
Voltage |
240 Vac +/-10%, 50Hz, 1A |
Size (HxWxD) |
150 x 215 x 260 cm |
Weight |
5 kg |
Calibration |
ISO 21501 - NIST traceable |
CCCN Code |
Tariff 90318020 |